Evaluation of Sysmex XE-2100 for enumeration and differentiation of cellular elements in peritoneal and pleural fluids

AUTORI

Sabrina Buoro1, Cosimo Ottomano2, Rosanna Gustinetti1, Claudia Galliani1, Eleonora Lochis1, Cheren Saleri1, Paola Dominoni1, Giuseppe Lippi2
1Laboratorio Analisi Chimico Cliniche, Azienda Ospedaliera Ospedali Riuniti, Bergamo
2Unità Operativa di Diagnostica Ematochimica, Azienda Ospedaliero-Universitaria di Parma

ABSTRACT

The aim of this study was to evaluate the performance of Sysmex XE-2100 for automated flow cytometric analysis of
biological fluids. The results of 106 consecutive peritoneal samples and 20 pleural samples were compared with optical microscopy. A good agreement was found in total nucleated cells (TNC) count between Sysmex XE-2100 and manual microscopy (y=0.98x+3.2, r=0.99; mean bias, -9.5 TNC/μL). The percentage of cells with high fluorescence in XE-2100 correlated with that of macrophages and mesothelial cells at microscopy (r=0.67; P <0.001). The CV of TNC determination on XE-2100 ranged between 1.6% and 11.7%. At the 20% CV, the analytical sensitivity of XE-2100 was 29/μL for TNC, 50/μL for polymorphonuclear leukocytes (PMN) and 66/μL for mononuclear elements. The linearity between 30 and 944 TNC/μL was excellent (r=0.998, P <0.001), whereas the carry-over was <1%. At thresholds of ≥250 PMN/μL in peritoneal fluid and ≥50% PMN or ≥50% lymphocytes in pleural fluid, the diagnostic accuracy of XE-2100 was 96.8%, with Cohen's kappa of 0.90 (P <0.001). We conclude that automated flow cytometric analysis on Sysmex XE-2100 may be a viable options for screening of peritoneal and pleural fluids.