Evaluation of a new anti-müllerian hormone (amh) assay: a comparison between atellica amh (Siemens Healthineers) and access amh (Beckman Coulter) assays

AUTORI

Antonino Sammartano1, Giulia Testa1, Lorenzo Barusi2, Luigi Ippolito1
1U.O. Clinical Pathology, Medical and Diagnostics Department P.O. Fidenza, AUSL Parma, Italy 2U.O. Obstetrics and Gynecology, Surgery Department P.O. Fidenza, AUSL Parma, Italy

ABSTRACT

Introduction: anti-müllerian Hormone (AMH) is a key marker of ovarian reserve; its serum levels are strongly correlated with the antral follicle count. Aim of this study is a comparison between two automated AMH assays: the new generation Atellica Solution® AMH sandwich immunoassay (AMH-AS) and the more widely used Access AMH immunochemiluminescence assay performed on Beckman-Coulter DxI analyzer (AMH-DxI) currently used in our laboratory. This evaluation has been performed according to the CLSI EP09-A3 guideline.
Methods: 120 consecutive serum samples with a AMH test request from the Medically Assisted Procreation center (PMA), have been used for the study. All samples were handled and processed in parallel. Statistical data analysis was performed using Stata program.
Results: passing-Bablok regression of AMH assays showed a good agreement between AMH-AS and AMH-DxI. The regression equation was AMH- AS = -0.0070+AMH-DxI x 0.994. In addition, method comparison did not show a large variance around the regression line. No significant differences neither in percentage or absolute values were identified through the Bland Altman test, thus showing a good concordance between the two methods.
Conclusions: method comparison is an important step in the validation process of assays and instruments. This comparison showed a good agreement between AMH-AS and AMH-DxI assays. Our results demonstrated an excellent analytical performance of fully automated AMH-AS immunoassay; therefore the use of this method will provide a better option in evaluating different clinical conditions including infertility treatment.

BIBLIOGRAFIA

  1. Taylor HS, Pla L, Seli E eds Speroff’s clinical Gynecologic Endocrinology and Infertility. Wolters Kluwer, Phyladelphia, PA 2020.
  2. Practice Committee of the American Society for Reproductive Medicine. Aging and infertility in women. Fertil Steril 2006;86 5 Suppl 1: S248-52.
  3. Practice Committee of the American Society for Reproductive Medicine. Electronic address: asrm@asrm.org; Practice Committee of the American Society for Reproductive Medicine. Testing and interpreting measures of ovarian reserve: a committee opinion. Fertil Steril 2020;114:1151-7.
  4. Dayal M, Sagar S, Chaurasia A, Singh U. Anti-mullerian hormone: a new marker of ovarian function. J Obstet Gynaecol India 2014;64:130-3.
  5. Moolhuijsen LME, Visser JA. Anti-Müllerian Hormone and Ovarian Reserve: Update on Assessing Ovarian Function. J Clin Endocrinol Metab 2020;105:3361-73.
  6. Gleicher N, Weghofer A, Barad DH. Discordances between follicle stimulating hormone (FSH) and anti-Müllerian hormone (AMH) in female infertility. Reprod Biol Endocrinol 2010;8:64.
  7. Iliodromiti S, Anderson RA, Nelson SM. Technical and performance characteristics of anti-Müllerian hormone and antral follicle count as biomarkers of ovarian response. Hum Reprod Update 2015;21:698-710.
  8. Pearson K, Long M, Prasad J, Wu YY, Bonifacio M. Assessment of the Access AMH assay as an automated, high-performance replacement for the AMH Generation II manual ELISA. Reprod Biol Endocrinol 2016;14:8.
  9. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;8476:307-10.
  10. Vidali M, Tronchin M, Dittadi R, per il Gruppo di Studio SIBioC – Medicina di Laboratorio “Statistica per il laboratorio. Protocol for the comparison of two laboratory methods. Biochim Clin 2016;40:129-42.
  11. Clinical and Laboratory Standards Institute (CLSI). Measurement procedure comparison and bias estimation using patient samples; Approved guideline. 3rd ed. CLSI document EP09. Wayne, PA, 2018.
  12. Dewailly D, Laven J. AMH as the primary marker for fertility. Eur J Endocrinol 2019;181:D45-51.
  13. Durlinger AL, Kramer P, Karels B, De Jong FH, Uilenbroek JT, Grootegoed JA et al. Control of primordial follicle recruitment by anti-Müllerian hormone in the mouse ovary. Endocrinology 1999;140:5789-96.
  14. Dewailly D, Andersen CY, Balen A, Broekmans F, Dilaver N, Fanchin R, et al. The physiology and clinical utility of anti-Mullerian hormone in women. Hum Reprod Update 2014;20:370-85.
  15. Ferguson J, Hockley J, Rigsby P, Burns C. Establishment of a WHO Reference Reagent for anti-Mullerian hormone. Reprod Biol Endocrinol 2020;18;86.
  16. Li HWR, Robertson DM, Burns C, Ledger WL. Challenges in Measuring AMH in the Clinical Setting. Front Endocrinol (Lausanne) 2021;12:691432.

HOME
PRIVACY POLICY
5x1000 Docemus

LOGO SIBioC

EDITORE RESPONSABILE
Alberto Oliaro

EDITORIAL SECRETARY
Edizioni Minerva Medica S.p.A.
Corso Bramante 83-85, 10126 Torino
T +39 011 678282
journals.dept@minervamedica.it

Designed by Biomedia srl
© 2025 SIBioC
P. IVA IT 06484860967