Modalità di refertazione del D-Dimero e prestazioni dei sistemi analitici: quali informazioni dal programma di Valutazione Esterna di Qualità del Centro di Ricerca Biomedica
Reporting of D-Dimer and performance of analytical systems: data from participants to the External Quality Assessment Scheme of the Biomedical Research Centre in Italy

AUTORI

Sandra Secchiero1, Laura Sciacovelli1, Mario Plebani1,2
1Centro di Ricerca Biomedica per la Qualità in Medicina di Laboratorio, Azienda Ospedale – Università, Padova, Italia
2Dipartimento di Medicina di Laboratorio, Azienda Ospedale – Università, Padova, Italia


Questo lavoro è stato in parte presentato al 52° Congresso Nazionale SIBioC, 6-8 Ottobre 2020, Virtual Edition, quale Comunicazione Orale

ABSTRACT

Reporting of D-Dimer and performance of analytical systems: data from participants to the External Quality Assessment Scheme of the Biomedical Research Centre in Italy

Introduction: D-Dimer assessment represents a cornerstone in the diagnostic approach to several thrombotic disorders. Recent literature has highlighted the role of D-Dimer also in the diagnostic pathway of coronavirus infection (COVID-19) and the importance of harmonized reporting [D-dimer unit (DDU) or fibrinogen equivalent unit (FEU); unit of measure; cut-off] in order to guarantee the correct interpretation of the results.
Methods: D-Dimer data from 100 EQA participants and the inter-laboratory variability (CV%) of the last 7 years for the most used analytical systems: Werfen, HemosIL HS; Werfen, HemosIL HS-500; Auto D-D, Sclavo; Innovance, Siemens; VIDAS, bioMérieux and STA Liatest, Stago have been evaluated.
Results: concerning the results expression in DDU or FEU, we observed a prevalence of FEU (55.1%) over DDU (44.9%); the value was confirmed in the last 7 years (average FEU = 55.6%), differently from data obtained in the survey conducted in 2014 at a national level.
The units used are: ng/mL (67.8%), μg/L (29.0%) and mg/L (3.2%) for D-Dimer DDU; ng/mL (57.9%), μg/mL (21.0%), μg/L (15.8%) and mg/L (5.3%) for D-Dimer FEU.
Inter-laboratory variability (mean CV%) calculated on a total of 72 controls is lower for all diagnostic systems at pathological levels than the one observed for concentrations below the cut-off.
Discussion: this study demonstrates that the reporting of D-Dimer results does not comply with the 2014 SIBioC consensus document which recommended the use of μg/L FEU, and highlights 8 different types of information. Data reported in this study call for the harmonization of D-Dimer reporting in order to guarantee the correct interpretation of the information, both in the case of COVID-19 and in all the diseases already known where this analyte has a clinical relevance.