• Passa al contenuto principale
BC

biochimica clinica

it_IT Italian
it_IT Italian en_US English
  • Home
  • Casi clinici
  • Ahead of print e Ultimo Fascicolo - Accedi per visualizzare gli articoli
  • Archivio BC fino a 2024
  • Sottometti un articolo
  • Norme Autori
  • Cerca

CONTRIBUTI SCIENTIFICI – Scientific Papers

Volume:

Biochimica Clinica 2012; 36(4) 240-243

Pubblicato on-line:

DOI:

Scarica in PDF:
Autenticazione richiesta

Preliminary evaluation of a new nephelometric reagent for the determination of serum free light chains

AUTORI

Fabio Duranti, Stefania Casciani, Paolo Casalino, Veronica Botteri, Maria Teresa Caliò, Ilio Giambini, Sergio Bernardini, Mariarita Dessi
Department of Laboratory Medicine, University Hospital “Tor Vergata”, Rome

ABSTRACT

The determination of free light chains (FLC) in serum of patients with monoclonal gammopathies is a valuable tool for diagnosis, follow-up, and prognosis of these conditions. The aim of this study was to compare the performance of the FLC reagents from New Scientific Company based on polyclonal antibodies with the Siemens assays (N Latex FLC) based on a mixture of monoclonal antibodies for the nephelometric determination of FLC in serum. Ninety-one serum samples were analyzed for FLC using both assays. Comparisons of the two FLC assays showed a slope of 4.87 (r=0.84) for ĸ FLC determination, a slope of 4.38 (r=0.83) for λ FLC determination, and a slope of 3.42 (r=0.60) for the k/λ FLC ratio. The concordance was 65% for ĸ FLC determinations, 63% for λ FLC, and 71% for k/λ FLC ratio. A preliminary evaluation suggests a higher sensitivity for N Latex FLC assays

BIBLIOGRAFIA

HOME
PRIVACY POLICY
5x1000 Docemus

LOGO SIBioC

EDITORE RESPONSABILE
Alberto Oliaro

EDITORIAL SECRETARY
Edizioni Minerva Medica S.p.A.
Corso Bramante 83-85, 10126 Torino
T +39 011 678282
journals.dept@minervamedica.it

Designed by Biomedia srl
© 2025 SIBioC
P. IVA IT 06484860967